Use of Flash Glucose-Sensing Technology for 12 months as a Replacement for Blood Glucose Monitoring in Insulin-treated Type 2 Diabetes.





Faculty/Professorship: Clinical Psychology and Psychotherapy  
Author(s): Haak, Thomas; Hanaire, Hélène; Ajjan, Ramzi; Hermanns, Norbert  ; Riveline, Jean-Pierre; Rayman, Gerry
Title of the Journal: Diabetes therapy : research, treatment and education of diabetes and related disorders
ISSN: 1869-6953
Publisher Information: New York, NY [u.a.] : Springer Healthcare Communications
Year of publication: 2017
Volume: 8
Issue: 3
Pages: 573-586
Language(s): English
DOI: 10.1007/s13300-017-0255-6
URL: https://link.springer.com/content/pdf/10.1007%2...
Abstract: 
INTRODUCTION: Published evaluations of sensor glucose monitoring use in insulin treated type 2 diabetes are limited. The aim of this study was to assess the impact of flash glucose-sensing technology as a replacement for self-monitoring of blood glucose (SMBG) over a 12-month period in participants with type 2 diabetes who were on intensive insulin therapy.
METHODS: An open-label, randomized, controlled study in adults with type 2 diabetes on intensive insulin therapy from 26 European diabetes centers aimed at assessing flash glucose sensing technology was conducted. Participants (N = 224) were randomized (1:2 respectively) to a control group (n = 75) that used SMBG (FreeStyle Lite™) or to an intervention group (n = 149) which used sensor glucose data (FreeStyle Libre™ Flash Glucose Monitoring System) for self-management over 6 months. All intervention group participants who completed the 6-month treatment phase continued into an additional 6-month open-access phase.
RESULTS: A total of 139 intervention participants completed the 6-month treatment phase and continued into the open-access phase. At 12 months (end of open-access period), time in hypoglycemia [sensor glucose <3.9 mmol/L (70 mg/dL)] was reduced by 50% compared to baseline [-0.70 ± 1.85/24 h (mean ± standard deviation); p = 0.0002]. Nocturnal hypoglycemia [2300 to 0600 hours, <3.9 mmol/L (70 mg/dL)] was reduced by 52%; p = 0.0002. There was no change in time in range [sensor glucose 3.9-10.0 mmol/L (70-180 mg/dL)]. SMBG testing fell from a mean of 3.9 (median 3.9) times/day at baseline to 0.2 (0.0), with an average frequency of sensor scanning of 7.1 (5.7) times/day at 12 months, and mean sensor utilization was 83.6 13.8% (median 88.3%) during the open-access phase. During this 6-month extension period no device-related serious adverse events were reported. Nine participants reported 16 instances of device-related adverse events (e.g. infection, allergy) and 28 participants (20.1%) experienced 134 occurrences of anticipated skin symptoms/sensor-insertion events expected with device use (e.g. erythema, itching and rash).
CONCLUSION: The use of flash glucose-sensing technology for glycemic management in individuals with type 2 diabetes treated by intensive insulin therapy over 12 months was associated with a sustained reduction in hypoglycemia and safely and effectively replaced SM
Keywords: Flash sensor glucose technology, Glucose monitoring, Insulin, Type 2 diabetes
Peer Reviewed: Ja
International Distribution: Ja
Open Access Journal: Ja
Type: Article
URI: https://fis.uni-bamberg.de/handle/uniba/43911
Year of publication: 22. June 2018